Perfusion Leadership & Management
Clinical ResearchIssue No. 3November 16, 20262 min read

The Marker That Stays Quiet

Washed red cells from pigs, cows, and people ran through one pump. All three shed vesicles. Only two of the three raised free hemoglobin. In the pig, the standard hemolysis marker went quiet while the damage was happening.

The Research

Foster KM, Foster M, El Banayosy AM, et al. Sensitivity of Damage Markers for Mechanical Trauma to Erythrocytes in Mechanical Circulatory Support Devices: A Multispecies Study. ASAIO Journal, 2026. doi:10.1097/MAT.0000000000002863. · PubMed 42912017

Washed red cells from pigs, cows, and people ran through one pump. All three shed vesicles. Only two of the three raised free hemoglobin. In the pig, the standard hemolysis marker went quiet while the damage was happening.

What They Did

Perfused washed red blood cells from three species through a CentriMag loop and measured two things over time: free hemoglobin, the marker everyone already watches, and erythrocyte-derived extracellular vesicles, which are what the membrane sheds when it is mechanically hurt. Same pump, same loop, three kinds of blood.

What They Found

All three species produced significant increases in vesicles. Only human and bovine cells produced significant increases in free hemoglobin. The normalized index of hemolysis did not differ between species, yet bovine cells generated far more vesicles than human or porcine, which were similar. The marker and the species were both doing work, and the standard marker read differently depending on which blood was in the loop.

The Hallway

The usual measure of red cell damage looked the same in all three species. The membrane the cells shed did not, and cow blood shed far more of it than human or pig.

Palm TreeThere were no differences in the normalized index of hemolysis between species, but bovine RBCs produced significantly more ErEVs than human or porcine, which were similar.

Where the Evidence Is Strong

The pump is held constant and only the blood changes, which is the clean way to ask whether the marker or the species is driving the answer. Both markers are measured in the same run, so the comparison is internal instead of borrowed from other papers. And it lands somewhere your own bench can test cheaply, because the question is simply whether the number you trust moves when the damage does.

Where the Evidence Is Limited

Washed cells in a loop are not blood in a patient. No plasma proteins, no endothelium, no complement, no heparin. Three species and one pump, so the CentriMag result may not carry to a different device. The authors say themselves that human sensitivity still needs testing, which means the more sensitive marker is not yet established in people. Vesicle measurement is not standardized across labs, so a count at one center may not mean the same at another. And more sensitive is not the same as clinically predictive, because nobody has shown that vesicle counts track outcomes.

The InFlow-mation

Critical Conclusion

Hemolysis alone under-reports mechanical damage to red cells, and in a porcine model it can under-report it to almost nothing. Vesicles moved in every species tested.

Take It Back to the Pump

A normal plasma free hemoglobin is not proof a circuit is gentle. When someone shows you pig data with a flat hemolysis index, ask what else they measured. For patients, keep this a research question and not a bedside test until somebody validates it in human blood.

Pump off.

On the source

Peer reviewed bench study with no patients. Washed red cells, three species, one centrifugal pump. No standardized vesicle assay exists and human sensitivity is untested, so do not carry this into a clinical protocol.